Purpose
This document describes the diode mode quantum efficiency
(QE) measurement process at the University of Arizona Imaging Technology
Laboratory. Diode mode QE testing is the process of measuring the photocurrent
generated from an entire detector (such as a CCD) and comparing it to the
photocurrent generated by a standard diode. The detector is not powered on for
this test. In the case of a CCD, the photocurrent is measured across the buried
channel, typically using the RD and SUB connections.
Equipment
- Software: National Instruments LabVIEW and Microsoft Excel
- Data acquisition board: National Instruments PCI-GPIB IEEE 488
- Photodiode: United Detector Technology, UV enhanced, PIN 10321, last
calibrated from 200 – 1200 nm on 8/20/2002
- Keithley Electrometer 6512
- For QE measurements with l ≥ 300 nm
(all Oriel Instruments part numbers):
Integrating sphere: Labsphere 20" diameter,
IAS-200 SF
Light source: 150W Xe
ozone free lamp (6255)
Arc lamp housing
(66002)
Universal power supply
(68805)
Light intensity
controller system (68850)
MultiSpec 257
monochromator (77702)
quadruple grating
turret (77708)
1 mm x 15 mm fixed
slits (77733)
Motorized filter
wheels (77737)
600 l/mm 200nm blaze
grating (77743)
600 l/mm ruled grating
(77744)
- For QE measurements with l < 300 nm
(all Oriel):
Integrating sphere: Labsphere 4" diameter,
US-040-SL CA-02117-000
Light source: 30W D2 ozone-free lamp (63161)
Deuterium lamp housing
(60023)
Deuterium lamp power
supply (68840)
Oriel interferences filters from 200 – 300 nm: 53310 (200 nm), 53320 (220 nm),
53330 (240 nm), 53340 (250 nm), 53350 (260 nm), 53355 (270 nm), 53360 (280 nm),
53365 (290 nm), 53370 (300 nm)
Measurement Overview
We place a calibrated diode in our optical beam and take
current readings at each wavelength of interest. The diode is then removed and
the detector (CCD) is located in the same position. Current readings are taken
from the detector at the same wavelengths. The detector may be in either a
dewar or in a basic mounting fixture. The values can be corrected for the
transmission/reflection loss of any dewar window and for quantum yield in the
UV.
A National Instrument’s LabVIEW program is used to acquire
and manipulate the current data. The program sets the monochromator to the starting wavelength and then acquires current readings from the
electrometer. The program scans over a specified range of wavelengths,
acquiring current readings for each wavelength. Typically, scans from 300nm to
400nm have increments of 20nm, and scans from 400nm to 1100nm have increments of
50nm. This is because there are typically rapid QE changes at shorter
wavelengths but relatively slow changes at longer wavelengths. The program
outputs the median of many current readings for each wavelength in the scan,
saving them to a spreadsheet file. This data is used to automatically compute
QE using an Excel program.
For l ≥ 300 nm, we
use an Oriel Hg-Xe 150 W short arc lamp as an illumination source. A
photo-feedback unit stabilizes the power supply driving the lamp. Because of
this stability (independence from line voltage variations) we do not measure the
calibrated photodiode after each wavelength setting but typically once per day.
While the short arc lamp produces more UV radiation than a
deuterium lamp, it is not as useful in the UV because of the large visible and
IR output. When making UV measurements (l < 300 nm), even a few tenths percent visible leak can swamp the UV intensity.
For this reason, we use a deuterium lamp for UV measurements. Because the flux
is very low in the UV when using our large integrating sphere, we have a
separate smaller integrating sphere and filter set for the 200 – 300 nm range.
Monochromator and filters
For l ≥ 300 nm,
we use an Oriel MultiSpec 257 monochrometer to select the desired wavelength for
each test. This unit is computer programmable via a RS232 interface. Two
filter wheels are controlled by the monochrometer for neutral density and order
blocking selection. The neutral density filters are required to reduce the
light intensity to ensure the signal on the CCD does not saturate the device.
An exit and entrance slit can be selected to specify a particular bandpass. We
typically use 1 mm slits providing about a 100 A bandpass with a 600 line
grating.
For l < 300 nm we
use an Oriel filter set and a dedicated, smaller integrating sphere. The basic
measurement process is the same, but the filter selection is controlled manually
rather than through the monochrometer.
Calibrated Diode
We use calibrated photodiodes as standards against which
the CCD QE is calculated. The diodes are UV enhanced and have 613 mm2 area. The diodes are placed in the same location as the CCD to minimize the
calibration requirements of the system. They are calibrated each year and
traceable to NIST standards.
Measurement Procedure for CCDs
-
Turn on optical system bench,
including all required lamps and the electrometer. Allow to warm up at least
15 minutes.
-
Place calibrated diode in
measurement fixture and connect to electrometer.
-
Make sure monochrometer is in
remote mode and run a diode scan using the QE Data Acquisition program.
-
Using the clean bench in the
Characterization Lab, place CCD in the appropriate fixture with at least Reset
Drain and Substrate wires connected to the BNC QE connector. Record the
serial number of the device.
-
Remove diode from test systems and
install CCD. Connect to electrometer.
-
Run a CCD scan.
-
Run Excel CCD-QE program. Select
device from pull down list to account for the light sensitive region of the
detector. Press Open Links to calculate QE from most recent scans.
-
Press Create QE sheet to store QE
data in a separate file. Save this new file as the ITL serial number (snxxxx.xls).
-
Remove device from test system and
uninstall from test fixture on clean bench.
-
Update ITL database for this
device.
|